Intrinsic Optical Properties of Boston Keratoprosthesis

Citation:

Hui P-C, Pereira LA, Dore R, Chen S, Taniguchi E, Chodosh J, Dohlman CH, Paschalis EI. Intrinsic Optical Properties of Boston Keratoprosthesis. Transl Vis Sci Technol 2020;9(12):10.

Date Published:

2020 Nov

Abstract:

Purpose: To benchmark the optical performance of Boston Keratoprosthesis (B-KPro). Methods: Back focal lengths (BFL) of B-KPros for various eye axial lengths were measured using an optical bench, International Organization for Standardization-certified for intraocular lens characterization, and compared against manufacturer's specification. The modulation transfer function (MTF) and the resolution efficiencies were measured. The theoretical geometry-dependent higher-order aberrations (HOA) were calculated. The devices were characterized with optical profilometry for estimating the surface scattering. Aberration correction and subsequent image quality improvement were simulated in CODE-V. Natural scene-imaging was performed in a mock ocular environment. Retrospective analysis of 15 B-KPro recipient eyes were presented to evaluate the possibility of achieving 20/20 best-corrected visual acuity (BCVA). Results: BFL measurements were in excellent agreement with the manufacturer-reported values (r = 0.999). The MTF specification exceeded what is required for achieving 20/20 visual acuity. Astigmatism and field curvature, correctable in simulations, were the primary aberrations limiting imaging performance. Profilometry of the anterior surface revealed nanoscale roughness (root-mean-square amplitude, 30-50 nm), contributing negligibly to optical scattering. Images of natural scenes obtained with a simulated B-KPro eye demonstrated good central vision, with 10/10 visual acuity (equivalent to 20/20). Full restoration of 20/20 BCVA was obtainable for over 9 years in some patients. Conclusions: Theoretical and experimental considerations demonstrate that B-KPro has the optical capacity to restore 20/20 BCVA in patients. Further image quality improvement can be anticipated through correction of HOAs. Translational Relevance: We establish an objective benchmark to characterize the optics of the B-KPro and other keratoprosthesis and propose design changes to allow improved vision in B-KPro patients.

Last updated on 11/30/2020