Comparison of swept-source and enhanced depth imaging spectral-domain optical coherence tomography in quantitative characterisation of the optic nerve head.

Citation:

Li D, Taniguchi EV, Cai S, Paschalis EI, Wang H, Miller JB, Turalba AV, Greenstein SH, Brauner S, Pasquale LR, Shen LQ. Comparison of swept-source and enhanced depth imaging spectral-domain optical coherence tomography in quantitative characterisation of the optic nerve head. Br J Ophthalmol 2016;

Date Published:

2016 Jun 13

Abstract:

AIMS: To compare swept-source optical coherence tomography (SS-OCT) and enhanced depth imaging spectral-domain OCT (EDI-OCT) in quantitative assessment of optic nerve head (ONH) parameters. METHODS: In a cross-sectional study, patients with primary open angle glaucoma (POAG) and age-matched control subjects underwent SS-OCT and EDI-OCT B-scans of the ONH in a single visit. Two masked readers independently measured the horizontal and vertical lamina cribrosa depth (LCDH and LCDV, respectively), as well as thinnest Bruch's membrane opening minimum rim width (BMO-MRW) from SS-OCT and EDI-OCT scans. We assessed agreement between SS-OCT and EDI-OCT measurements by linear regression models, Bland-Altman analysis and concordance correlation coefficients (CCC). Intrareader and inter-reader reproducibility was assessed using intraclass correlation coefficients (ICC). RESULTS: One eye from each of 40 patients with POAG and 20 controls were included. All three ONH measurements were higher on SS-OCT than on EDI-OCT, with significant differences in LCDH (mean difference=31.7 µm, p<0.01) and thinnest BMO-MRW (mean difference=20.5 µm, p<0.01). Linear regression models described the agreement between SS-OCT and EDI-OCT measurements with R(2)>0.8 for LCDH among both patients with POAG and controls and for thinnest BMO-MRW among patients with POAG. The CCC was >0.8 overall for each parameter. Intrareader and inter-reader ICCs were ≥0.989 and ≥0.964, respectively, for all parameters. CONCLUSIONS: LCDH, LCDV and thinnest BMO-MRW measurements are not interchangeable between SS-OCT and EDI-OCT, but show good intrareader and inter-reader reproducibility and interdevice agreement for quantitative characterisation of the ONH, particularly among patients with glaucoma.

See also: June 2016, All, 2016
Last updated on 06/24/2016