UNLABELLED: We present a case of corneal perforation secondary to an intrastromal astigmatic keratotomy performed with an optical coherence tomography-guided femtosecond laser. The keratotomy was concomitant with cataract surgery and resulted in a flat anterior chamber prior to the start of lens extraction. Interrupted nylon sutures were placed to seal the keratotomy prior to phacoemulsification. Escape of cavitation bubbles into the anterior chamber or the liquid interface can alert the surgeon to the possibility of unintended perforation of the endothelium or the epithelium, respectively. FINANCIAL DISCLOSURE: Neither author has a financial or proprietary interest in any material or method mentioned.